Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

Proposed enhancementBrief descriptionTODOsRelevant jira ticket (s)Meeting links where this was discussedStatusPriorityUpdate date
TSO3 Outlier Detection

Currently, the JWST pipeline does not properly use TSO information to reject outliers (e.g., jump step or current tso3 outlier rejection — see Jira tickets). Proposed to implement HST/STIS algorithm developed by Nikolay Nikolov on JWST TSO pipeline. 

  • Algorithm currently tried only on NIRCam data. Need to try it on NIRISS/SOSS, MIRI/LRS and NIRSpec/BOTs.

Problems with current TSO3 step:

Jira
serverSTScI JIRA
serverIdaac911eb-9c55-3f14-8368-77b436fb6dc2
keyJP-1285
Jira
serverSTScI JIRA
serverIdaac911eb-9c55-3f14-8368-77b436fb6dc2
keyJP-1654
Jira
serverSTScI JIRA
serverIdaac911eb-9c55-3f14-8368-77b436fb6dc2
keyJP-1647

CalWebb WG: 2021-05-04 Meeting Notes; 2021-01-05 Meeting Notes.

TSO WG: 2021-02-24 TSO WG Meeting notes.

Status
colourYellow
titleProposed

3December 2, 2021
Jump detection using TSO information

Similar to the above, but do it at the group-level; detect jumps using TSO information. This is an idea proposed by Michael Regan. Pros: you don't rely on the reference files.

  • Need to do simulations to test algorithm.


Status
titleIDEA

2
Correct jitter spectral movement prior to photom stepIf jitter is an issue during JWST observations, then the photom step will incorrectly apply photometric flux standarization to the wrong wavelengths, introducing time-dependant systematics. TSO WG reps think turning off the photom step would be ideal, but not generic solution. Correcting those jitters prior to the photom step would be a solution.
  • Need to write cross-correlation function to correct spectra.
  • Apply it to spectra; and see if this can correct photom step jitter issues.

Jira
serverSTScI JIRA
serverIdaac911eb-9c55-3f14-8368-77b436fb6dc2
keyJP-2082

CalWebb WG: 2021-06-01 Meeting Notes.

TSO WG: 2021-06-02 TSO WG Meeting notes.

Status
titleIDEA

2
Spectral tracingPipeline needs to perform spectral tracing of TSOs in order to (a) record trace movements (useful as external parameters to decorrelate lightcurves) and (b) extract the same portion of the spectrum at each time.
  • Define tracing algorithms for each instrument.
  • Test those.
  • Implement on the pipeline.


Status
titleIDEA

2December 15, 2021
Photometric CentroidingSame as spectral tracing, but for photometry.



2
Pre-amp reset correction ("reference pixels for subarrays without reference pixels")
  • Pipeline needs to take care of detector effects not taken care of by, e.g., superbias in the case where there are no reference pixels.
  • Everett Schlawin proposes a "fast-by-fast", "slow-by-slow" substraction; give pipeline pixels that it can use on each mode (= "unilluminated pixels").
  • Algorithms like these are useful not only for subarrays with reference pixels: using noniluminated pixels is even better than "just" using reference pixels.
  • Important for both spectroscopy and photometry


Status
titleIDEA

1January 3, 2022
Background substraction 
  • Background substraction is embedded on extract1d. But this is unintuitive.
  • Need to allow for 2D background profile (e.g., we know NIRISS/SOSS has this!).
  • This is particularly troublesome for MIRI/LRS. 



Status
titleIDEA

2January 3, 2022
Optimal extraction
  • Pipeline currently does box-extractions. Should perform extraction by weighting pixels by their SNRs/noise levels.

  • Optimal extraction could include likelihood information on 1/f noise, for instance.


Status
titleIDEA

2January 3, 2022
Pixel timing
  • Due to the sequential nature of the pixel readout modes, every pixel does not have the same time-stamp. The pipeline should either report this, or have an utility function that corrects for this.
  • Timing of the pixels is known, but uncertainties are not.

Discussion on this:

Jira
serverSTScI JIRA
serverIdaac911eb-9c55-3f14-8368-77b436fb6dc2
keyJP-2330


Status
titleIDEA

1December 15, 2021
Time-stamp accuracy
  • Pipeline reports times in BJD — but are uncertainties associated with that? What is the precision on this? Has this been tested? (spacecraft position is known accurately probably, but this has a limit? )

  • Important to note is that there are some Cycle 1 programs aiming to perform in-exposure timing measurements, but not inter-exposure accuracy measurements.



Status
titleIDEA

1
Integration-level aperture positions
  • Both for photometry and spectroscopy, a TSO aperture position is needed to be defined at each integration, because jitter might make the PSF/spectrum to change in position as a function of time (e.g., due to pointing errors, drifts, etc.).

  • This can be "hacked" currently, but its very messy to do.



Status
titleIDEA

3