Page History
...
Stage 2: Spectroscopic processing TSO testing
Pipeline step | NIRISS status | NIRISS comments | NIRCam status | NIRCam comments | MIRI status | MIRI comments | NIRSpec status | NIRSpec comments | ||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
assign_wcs |
|
|
|
| ||||||||||||||||||||||||||||||||
extract_2d |
|
|
|
| ||||||||||||||||||||||||||||||||
flat_field |
|
|
|
Build 7.5: Failed current validation test | passes for NRS2, but fails for NRS1: JP-1071 -NIRSpec: flat_field calspec2 step failing validation for FS ALLSLITS, BOTS, and MOS data OPEN | |||||||||||||||||||||||||||||||
srctype |
|
|
|
Build 7.5: Logic is not correct, always defaults to POINT, even when provided EXTENDED | JP-1576 - source_type logic in spec2 is not correct for NIRSpec FS and MOS modes OPEN | |||||||||||||||||||||||||||||||
photom |
|
|
|
| ||||||||||||||||||||||||||||||||
extract_1d |
| This step currently extracts the whole subarray as a spectrum, which produced order overlap, which makes this step produce an unusable product for science. An algorithm is in development for this step. For details, see
|
|
|
Build 7.5: methodology needs to be vetted by TSOWG | additional comments pending |
Extended comments by instrument branch:
...
Pipeline step | NIRISS status | NIRISS comments | NIRCam status | NIRCam comments | MIRI status | MIRI comments | NIRSpec status | NIRSpec comments | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
outlier_detection |
|
|
|
| ||||||||||||||||||||
tso_photometry |
|
|
|
| ||||||||||||||||||||
extract_1d |
|
|
|
| ||||||||||||||||||||
white_light |
|
|
|
|
Extended comments by instrument branch:
...